16KPa Pressure 2mL Synthetic Blood Penetration Tester Time Accuracy 0.1s

MOQ:1
Price:Negotiable
StandardPackaging:Plywood
DeliveryPeriod:30Days
PaymentMethod:T/T,L/C,WesternUnion,MoneyGram
SupplyCapacity:20SetsPerMonth

Products Description

Place of Origin:China

Brand Name:KingPo

Certification:Calibration Certificate (Cost Additional)

Model Number:KP122606

Test Pressure:0~16KPa

Micro-pressure Measuring Range:0~0.1KPa

Pressure Value Resolution:0.1kPa

Time Accuracy:0.1s

Test Medium:Synthetic Blood

Blood Volume:2ml

Minimum Order Quantity:1

Price:Negotiable

Packaging Details:Plywood

Delivery Time:30 Days

Payment Terms:T/T, L/C, Western Union, MoneyGram

Supply Ability:20 Sets Per Month

Highlight:mask testing machine, mask testing equipment


Product Description

 

Medical Masks 16KPa Pressure 2mL Synthetic Blood Penetration Tester

 

Product overview:

The medical mask synthetic blood penetration tester is used to measure the jet penetration performance of liquid sources such as synthetic blood to evaluate mask’s safety protection performance. Mainly used to test the permeability of masks to blood, body fluids, blood pathogens (tested with Phi-X174 antibiotics), synthetic blood and other liquids. It is suitable for liquid penetration resistance of protective equipment such as masks, gloves, protective clothing, outer cover, coveralls, boots and other protective equipment. This equipment can output test report and save test data.

 

Test method:

 

The medical mask sample is supported on the test device, and 2mL of synthetic blood with a surface tension of 0.042 ± 0.002n / m is sprayed from a cannula with an inner diameter of 0.84mm at a distance of 300mm to the mask under test at a pressure of 16kpa in the horizontal direction. Simulate the scene of the mask being splashed by blood from the perforated blood vessel. Take off and visually check the penetration of synthetic blood on the other side of the mask within 10 seconds. Test the medical mask with the corresponding liquid jet pressure of 450cm / s, 550cm / s, 635cm / s, record the test result of each jet velocity, give out the highest blood pressure corresponding to the acceptable quality level limit of 4 for medical masks.

 

Technical parameters:

 

Basic test principle Simulate breathing through a vacuum pump, and carry out the pressure difference detection
Main parameters Test pressure 0~16KPa
Micro-pressure measuring range 0~0.1KPa
Pressure value resolution 0.1kPa
Time accuracy 0.1s
Test medium Synthetic blood
Control method Manual
Blood volume 2ml
Scope of application Surgical mask
 
 