UL507 Fig 9.2 Stainless Steel Straight Unjoint Enclosure Protection Test Probe

MOQ:1
Price:Customized
StandardPackaging:AluminiumBox
DeliveryPeriod:15Days
PaymentMethod:T/T
SupplyCapacity:50setspermonth

Products Description

Place of Origin:China

Brand Name:KingPo

Certification:Calibration Certificate (Cost Additional)

Model Number:KP2210-35S

Applied Standard:UL507, UL474 And Etc.

Applied Figture:Fig 9.2 And Etc

Probe Material:Stainless Steel

Handle Material:Nylon

Applied Force:Should Work With Force Gauge

Package Type:Aluminum Case

Brand Name:KingPo

Warranty Time:One Year

Minimum Order Quantity:1

Price:Customized

Packaging Details:Aluminium Box

Delivery Time:15 Days

Payment Terms:T/T

Supply Ability:50 sets per month

Highlight:Straight Unjoint Enclosure Protection Test Probe, Stainless Steel Enclosure Protection Test Probe, UL507 Enclosure Protection Test Probe


Product Description

 

UL507 Fig 9.2 Stainless Steel Straight Unjoint Enclosure Protection Test Probe

 

Product overview:

The Joint Test Probe of Figure meets the standard requirements of UL507 Fig 9.2 (PA100A), IEC 62368-1:2018 clause 4.8, 5.3.2, 8.5.4.3, V.1.2 and figure V.1, etc., UL474 Fig 5.1, UL982 Fig 7.1, UL1017 Fig 2, UL1062 Fig1.2, UL1278Fig 8.3, UL1310 Fig 16.2, UL6500 Fig14, UL60065 Fig14 and UL60950 Fig NAF.1.

 

Technical Parameters:

Probe material Stainless steel
Handle material Nylon
Applied force Without built-in force, should work with a force gauge
Applied standards UL507 Fig 9.2 (PA100A), IEC 62368-1:2018 and etc.

 

Straight unjointed version                                           Jointed version

 

Test procedures and operation

Please refer to the relevant standards.

Take UL 474-2012 clause 5.2 as an example:

An opening is acceptable if the probe illustrated in Figure 5.1 cannot contact uninsulated live parts film coated wire, moving parts, and heater elements. The probe is to be applied:

a) With a force of 2.5 pounds (11.1 N) and

b) In any possible configuration and to any depth that the size of an opening will permit.

The probe is to be rotated or angled to any possible position before, during, or after insertion through the opening; and, if necessary, the configuration of the probe may be changed after the probe has been inserted through the opening.

 

The test probe doesn’t have built-in force, if specific applied force required, the test probe should work together with a force gauge. At the end of our probe handle, there is a M6 thread hole to match the force gauge. Showing as following reference picture:

 

Notice

1. Please do not damage the appliance shape (such as scratching, bending, etc.).

2. Prevented from dust and moisture environment, in order to avoid the oxidation to affect the dimensional accuracy.

Tags:

Related Products