Products
IEC 62133-1 Battery Vibration Impact Test Bench With Real Time Display Function
Products Description
IEC 62133-1 Battery Vibration Impact Test Bench With Real-Time Display Function
Overview:
This vibration system conforms to IEC 60068-2-6, it is required by IEC 62133-1:2017 clause 7.2.2, IEC 62133-2:2017 clause 7.3.8.1,etc.
Technical Parameters:
1.Sinusoidal Control Measurement System Specifications and Functions:
Setup function: The test specification of the human-machine dialogue setup can be saved for reuse
The controller can be on the rack, integrated with the amplifier cabinet, and can also be separated from the amplifier cabinet
2. Random Control Measurement System Specifications and Functions:
3.Impact control measurement system technology standard and function:
4. Characteristics and Technical Specifications of USB01 Card:
Vibration generator
Sinusoidal thrust
2940N (300kg.f)
Random thrust
2940N (300kg.f)
Impact thrust
5880N(600kg.f)
Frequency range
DG~3500Hz
Continuous displacement
20mm
Impact displacement (maximum displacement)
25mm
Maximum speed
1.8m/s
Maximum acceleration
981m/s2 (100g) (No-load)
Dynamic diameter
Φ150mm
First-order resonant frequency
2900 Hz±5%
Allowable eccentric moment
≤2500N·cm
Equivalent mass of moving parts
3kg
Load connection point
13
Table screw size (standard)
M8
Table screw layout (diameter, Yuan Zhou)
13-M8 Depth 16mm
Axial vibration isolation frequency
<3Hz
Maximum load
120kg
Magnetic flux leakage
≤1mT
Ambient temperature
0~40℃
Size
1600mm×800mm×700mm
Weight
1100kg
Power amplifier output
3kVA
Signal-to-noise ratio
>65dB
System protection
Rated output voltage
100Vrms
Output current
With the 150A step, the maximum can be up to 6000A
Input impedance
≥15kΩ
Input signal voltage
≤1.5Rms
Efficiency of power amplifier
>92%
Harmonic distortion (resistance load)
<1.0%
Output voltage measurement error
≤5%
Output current measurement error
≤5%
Output current crest factor
≥3
DC stability
Zero drift at the output is less than 30mv/8h
Frequency range
1-3500Hz:±3dB ; Medium frequency gain:≥80
DC/AC conversion efficiency
>92%
Load property
Resistive, capacitive, emotionally arbitrary
Unbalance of parallel flow
≤2.8%
Average time free working time(MTBF)
>3500 hours
Cooling mode
Forced air cooling
Power
AC 3phase 380V±10%, 60Hz
Size
600mm×620mm×1600mm
Weight
580kg
Cooling fan with muffler
Fan power
0.75kW
Air volume
0.1m3/s
Duct diameter
125mm
Duct length
4m
wind pressure
1kpa
Size
500mm×520mm×650mm
Weight
30kg
Horizontal extension slide
Platform size
500×500mm
Equivalent mass (maximum load-bearing)
28kg
Upper bound frequency
Sine 2000Hz, random 2000HZ
Material
Magnesium aluminum alloy
Vertical Expansion Table
Platform size
500mm×500mm
Equivalent mass (maximum load-bearing)
28kg
Upper bound frequency
Sine 1500Hz,random 2000HZ
Material
Magnesium aluminum alloy
Frequency range
2-5000Hz
Crossing point
32 segments, and can wait for displacement, equal speed, etc.
Frequency display
The vibration test curve set by the man-machine dialogue can be directly displayed on the screen. In the vibration test, the vibration curve can be displayed in real time, and the error dB at the time can be displayed continuously.
Output voltage
7Vrms
Sweep mode
Linear, logarithmic
Sweep rate
0.1-10oct/min
Control accuracy
±0.5dB
Dynamic Range
79dB
Control method
Peak, average, rms, tracking filter
Vibration Mode
Constant Frequency, Sweep (Linear, Logarithmic), Pause, Resume, Return
Filter form
Biased anti-aliasing filter (eight-stage) passband 0.2dB/oct
Signal form
Sine wave
Waveform distortion
≤0.5%
Frequency range
100,200,400,500,1000,2000,4000,5000Hz
Resolution
100,200,400,800 lines
Driving spectrum dynamic range
60dB
Control spectrum dynamic range
40dB
Loop time
2000Hz, 400 lines single channel 0.6 seconds
Power spectrum control accuracy
±1dB (90% confidence)
RMS value control accuracy
±0.2dB
Control Channels
Independent Channels
Control method
Average, maximum
Spectral settings
Up to 32 levels, both with PSD [(m.s2)2/Hz] and dB/oct settings
Waveform
Half Sine, Triangle, Sawtooth, Square Wave
Frequency Range
5-2000Hz
Vibration level presets
-18dB, -15dB, -12dB,-9dB,-6dB,-3dB,0dB seven vibration levels, vibration time for each vibration level :Set by the user.
Upgrade Modes
Automatic upgrade and manual upgrade
Display function
With real-time display, print function.
Protection function
Over-limit alarm, open-circuit protection and automatic over-limit shutdown
Setting function
The test specification set during the man-machine conversation can be saved for repeated use.
Measuring range
0-200g divided into eight gears 1,2.5,5,10,25,50,100,200g range automatically shift
Filter Format
Programmable anti-aliasing filter (eight-stage) 0.2dB pass band and -55dB/oct stop band
Signal form
True random, pseudo-random
Full-scale input
16-bit 16-channel single-ended A/D, speed 400KS/S, +/-5V
Full-scale output
One 16-bit D/A, +/-5V
Win-XP
Supports USB1.1 and USB2.0 standards for Win-2000
Main Controller
32-bit 120MHz Floating Point Digital Signal Processor (DSP)
Digital output
8 bits
High-speed continuous acquisition
34K*32Bit high-speed static memory
Clock signal output
819.2K filter
Input Channel
16-channel single-ended
Analog Input Signal Voltage Range
Bipolar +/-5V
Resolution
16 bits
A/D conversion time
2.5 microseconds
Acquisition pass rate
400KHz
Accuracy
Better than 0.015%
Nonlinear deviation
<4 LSB
Data Format
16-bit twos complement
Input impedance
10 MΩ
A/D conversion trigger mode
Instruction, timer
D/A output signal voltage range
Bipolar +/-5V
D/A resolution
16 bits
Digital/analog conversion time
5 microseconds
Accuracy
better than 0.015%
Nonlinear deviation
<4 LSB
Data Format
16-bit twos complement
Output impedance
<0.1Ω
Main controller
32-bit 120MHz floating point DSP TMS320VC33
System clock
40.96KHz
Two 32-bit timer/counters
For D/A and A/D timing
Digital output
8 bits(TTL)
Clock output
819.2KHz
Power Consumption
<2.5W(5V, 500mA)