ISO 5151 60K BTU Computer Control Dual Code Air - Enthalpy Calorimeter

MOQ:1
Price:Customized
StandardPackaging:Plywood
DeliveryPeriod:3Months
PaymentMethod:L/C,T/T
SupplyCapacity:1SetsPerMonth

Products Description

Place of Origin:China

Brand Name:KingPo

Certification:Calibration Certificate (Cost Additional)

Model Number:KP331C-60KD

Single Phase Voltage:0~300V

Single Phase Power:0~5kW

Single Phase Current:0~20A

Frequency:50/60Hz

Power Factor:0~1

Air Flow B:200~2000 M³/h

Air Flow A:300~4000 M³/h

B Station Heating Capacity:6000~30000Btu/h

Minimum Order Quantity:1

Price:Customized

Packaging Details:Plywood

Delivery Time:3 Months

Payment Terms:L/C, T/T

Supply Ability:1 Sets Per Month

Highlight:Dual Code Air Enthalpy Calorimeter, BTU Computer Control Air Enthalpy Calorimeter


Product Description

 

ISO 5151 60K BTU Computer Control Dual Code Air-Enthalpy Calorimeter

 

System introduction:

 

The air enthalpy calorimeter test chamber consists of an indoor side chamber (A chamber) and an outdoor side chamber (B chamber). The refrigerating capacity, heating capacity and other performance of the air conditioner are tested by testing the parameters of temperature, pressure and electrical performance of the laboratory. The device can be used to test the performance of various window type, split type, independent type, ceiling box type and duct type air conditioners, and can also be used for related tests and extended tests under other standard operating conditions to meet the needs of product sampling inspection and product development. The laboratory has two stations, can simultaneously test two sets of not more than 30KBtu single-phase prototype (to ensure that the total is not greater than 60K BTU), if only test a prototype once, the prototype can reach 60K BTU, but should not be greater than 60K BTU.

 

Testing items:

 

Rated cooling capacity (T1, T2, T3).

Rated heating capacity (high, low, ultra low)

Airflow

Dehumidifying capacity

Input power

► EER(COP)

Frequency

electric current

Voltage

 

Types of UUT and environment:

 

Unit under test (UUT) type

- Split type

- Windowed

- Freestanding

- Ceiling Box

- Duct type

- Inverter air conditioner

 

► Measurement range of UUT

- Test Capability Range:

- Total cooling capacity: 6000 ~ 60000 Btu / h

Cooling capacity of station A: 6000 ~ 60000 Btu / h

Cooling capacity of station B: 6000 ~ 30000 Btu / h

- Heating capacity: 6000 ~ 60000 Btu / h

Heating capacity of station A: 6000 ~ 60000 Btu / h

 

- Airflow Range

- Air flow rate A: 300 ~ 4000 m3 / h

Air volume B: 200 ~ 2000 m3 / h

- Overall Dimensions: Window: Width: Up to 1000 mm Depth: Up to 800 mm Height: Up to 1000 mm

Split type: width: less than 1000 mm depth: less than 500 mm height: less than 2000 mm

- Electrical parameters

- Single-phase power: 0 ~ 5 kW

Single phase voltage: 0 ~ 300 V

Single phase current: 0 ~ 20 A

Frequency: 50 / 60 Hz

- Power factor: 0 ~ 1

 

Major test conditions

No

Testing Item

Indoor side(℃)

Outdoor side(℃)

Dry Bulb

Wet Bulb

Dry Bulb

Wet Bulb

1

Rating Cooling

T1

27

19

35

24

T2

21

15

27

19

T3

29

19

46

24

2

Maximum Cooling

T1

32

23

48

32

T2

27

19

35

24

T3

32

23

52

31

3

Freeze-up

T1

21

15

21

15

T2

21

15

10

8

T3

21

15

21

15

4

Minimum Cooling

21

15

18

12

5

Enclosure Sweat

27

24

27

24

6

Rating Heating

High

20

15

7

6

Low

20

15

2

1

Extra Low

10

8

-5

-6

7

Maximum Heating

27

19

24

18

8

Minimum Heating

20

15

-7

-8

9

Automatic Defrost

20

12

2

1

 

KingPo is a diversified high-tech enterprise integrating testing equipment, scientific research and teaching equipment, all kinds of non-standard customized equipment, and integrating independent R & D design, production and sales, installation and commissioning, after-sales maintenance. Company technical staff has accumulated rich practical experience.

 

Tags:

Related Products