Products Description
Face Mask Respiration Resistance Testing Equipment
Respiration Resistance Face Mask Tester ±1000Pa 85±1L/ Min With Standard Test Head Model
Product overview:
The mask breathing resistance detector is used to measure the inhalation and exhalation resistance of the mask under specified conditions. It is applicable to carry out relevant testing and inspection of mask products for mask manufacturers and national labor protection products inspection agencies.
Equipment Features:
Test Principle
Meet GB 2626-2019 Respiratory protection Self-priming filter anti-particulate respirator.
2. Mechanical flow meter with high flow control accuracy;
3. Automatic constant current control, automatic sample qualification judgment;
4. Can set two modes of exhalation detection and inhalation detection;
5. Large-capacity data storage, real-time storage of test data;
6. Export historical data through U disk or thermal printer;
7. Parameters such as qualified judgment pressure difference and sample number can be set;
8. Automatic protection for fault detection.
Main Parameters
Range
Resolution
Max. Allowed Error
Test Flow
(85±1)L/min
0.1L/min
±1%
Pressure Range
±1000Pa
0.1Pa
±1%
Data Storage
100 groups
Working Power Supply
AC 220V±10% 50Hz or match client’s local power supply requirements
External Dimensions
420×600×720(L×W×H)
Equipment Weight
Approx. 30kg
Power